Drager Chlorine Dioxide 0-20ppm Sensor

PRODUCT CODE:
6811360
Drager Chlorine Dioxide 0-20ppm Sensor

Please note this product is currently discontinued.

Loading...

Product description

Drager Chlorine Dioxide 0-20ppm Sensor

This DragerSensor XS EC is suitable for the detection of Chlorine Dioxide (ClO2).  This particular sensor is compatible with the X-am 7000 multi gas detector and can be used in industries and segments such as food and beverage, breweries, wastewater treatment, swimming pools, industrial gases, pulp and paper.

It has a measuring range of 0 to 20 ppm.

Key Features

  • Plug and play sensor
  • Comes with a 1 year guarantee
  • Is especially selective
  • Features a particularly low cross sensitivity to chlorine
Specification

Detection Limit

0.02 ppm

Resolution

0.01 ppm

Response Time

≤ 20 seconds at 20 °C or 68 °F (T50)

Measurement accuracy

Zero Point:  ≤ ± 0.02 ppm
Sensitivity:  ≤ ± 5% of measured value

Long-term drift, at 20°C (68°F)

Zero Point:  ≤ ± 0.03 ppm/year
Sensitivity:  ≤ ± 2% of measured value/month

Warm-up time:

≤ 12 hours

Ambient conditions

Temperature: (–20 to 50)°C (–4 to 122)°F
Humidity: (10 to 90)% RH
Pressure: (700 to 1,300) hPa

Influence of Temperature

Zero Point:  ≤ ± 0.02 ppm
Sensitivity:  ≤ ± 5% of measured value

Influence of Humidity

Zero Point:  No Effect
Sensitivity:  ≤ ± 0.1% of measured value/% RH

Test Gas

ClO2 test gas between 40% and 100% of highest figure in the set measuring range. Every time this sensor is used, a function test should be done beforehand, for which the ClO2 generator (68 11 497) can be used.