Drager Diagnostics Kit (includes DDT5000/Alcotest 6820)
PLEASE NOTE: THIS KIT IS NO LONGER AVAILABLE. ITEMS MUST BE PURCHASED SEPARATELY.
This Diagnostics Kit from Drager contains Drager Drug Tester 5000 analyser unit, Printer, Carry case for the DDT5000, 6-panel test kit, Alcotest 6820 unit & 100 mouthpieces.
The Drager DrugTest® 5000 system is a fast, accurate means of testing oral fluid samples for drugs of abuse, such as amphetamines, designer amphetamines, opiates, cocaine and metabolites, benzodiazepines and cannabinoids, as well as precise diagnostic evaluation and data management.
The Drager DrugTest® 5000 system is comprised of two main components: the Drager DrugTest® 5000 Test Kits and the Drager DrugTest® 5000 Analyzer. Reliable and easy to use, the Dräger DrugTest®
5000 Analyzer is a mobile oral fluid screening device for providing qualitative detection of drugs using the Dräger DrugTest® 5000 Test-kits.
Providing reliable and precise analysis of the Test Kit sample within just a few minutes, this state-of-the-art opto-electronic system weighs less than 10 pounds. The Dräger DrugTest® 5000 Analyzer has a user-friendly, menu-driven, illuminated display which can be easily read, even at acute angles.
The Analyzer interprets the results from the DrugTest® 5000 Kit, eliminating any possibility of human error in reading the results. The full color display interface either indicates “Positive” or “Negative” for each drug class.
The Drager DrugTest® 5000 Analyzer, with its simple three-key operation, can also be easily linked to a wide variety of data recording devices such as a PC, Dräger Mobile Printer or barcode scanner.
The Analyzer’s integral data memory allows for up to 500 individual measurements to be stored for printing or upload to a PC.
In addition to data management, a built-in self-test capability controls temperature, optics and general operation.
With easily configurable software, the flexible menu can also be quickly configured to meet the needs of different applications.
Combined with a carry bag, this mobile system is a complete “substance abuse monitoring” system for on-the-spot measurements when equipped with a Drager DrugTest® 5000 Analyzer, Drager Mobile
Printer, keyboard, DrugTest® 5000 Test Kits, breath alcohol testing device, mouthpieces, and system documentation.
For additional information on the Alcotest 6820, please click here
Drug Test 5000
Supply Voltage | 12 V DC (11 to 15 V DC) |
Current Consumption | Typically 3 A |
Dimensions (W x H x D) | 7.9” x 9.8” x 8.7 |
Weight | 9.9 lbs |
Temperature | Operation: 5 to 40 °C (+41 to +104 °F) Storage/Transportation: -20 to +60 °C (-4 to +140 °F) |
Moisture Range | 5 to 95 %rh, non condensing |
Interfaces | Optical IRDA (printer), PS/2, USB Slave |
Duration of One Measurement | 5 to 10 minutes, depending on type of test |
Storage Capacity | 500 data sets |
Carrying Bag | with Complete System |
Alcotest 6820
Test Method | Electrochemical Dräger sensor for 1/4“ technology; alcohol-specific |
Test Range | 0 to 2.5 mg/L; display indication when exceeding test range |
Sampling | Standard: automatic sampling when minimum volume is reached Passive sampling or manual triggering of sampling possible |
Ready For Use | Approx. 2 s after powering up |
Test Result Display | Approx. 3 s (with 0 mg/l), 10 s (at 0.5 mg/l, room temperature) |
Operating Temperature | -5 to +50 °C; -4 to +122 °F |
Display | Graphic LCD display with background illumination; 41 mm x 24 mm (128 x 64 pixels) |
LED | 3 colours to indicate messages or alerts |
Acoustic Alarms | Different signal tones indicate various messages and alerts |
Memory | Space for saving the last 5000 tests with test number, date and time |
Power Supply | Two AA batteries or NiMH batteries, battery indicator on the display. Around 1500 breath tests can be conducted with one set of batteries. Charging rechargeable NiMH batteries is possible in device |
Calibration | Wet gas or dry gas calibration |
Dimensions (H x W x D), weight |
151 mm x 62 mm x 38 mm, approx. 195 g incl. batteries |
Device configuration | Device settings configurable directly via device menu (PIN required). No additional PC software required |
Vibration and Impact | Meets EN 60068-2-6, EN 60068-2-29 |
CE Marking | Directive 89/336/EEC (electromagnetic compatibility) |
Data Interface | USB to PC; optical interface for printer control |
Standard | Meets EN 15964 |
Protection Type | IP54 |
Printer
Operating Temperature | 23 °F to 113 °F |
Storage Temperature | -4 °F to 122 °F |
Humidity | 100%, not condensing |
Atmospheric Pressure | 600 to 1050 mbar |
Dimensions | 4.3” x 2.4” x 8.1” (W x H x D in inches) |
Weight | 1 lb |
Power Supply | 4 x 1.5V alkaline (AA) batteries |