Drager X-am 5000 Gas Detection Kit CatEx/O2/CO&H2S/Cl2

PRODUCT CODE:
3360222
Drager X-am 5000 Gas Detection Kit CatEx/O2/CO&H2S/Cl2
£3,603.04 inc VAT
£3,002.53

Additional services

  • £50.00
Gas detection equipment requires service and calibration every 6 months. In acknowledgement of this we are offering you your first calibration at a special rate.
Loading...

Product description

Drager X-am 5000 Gas Detection Kit CatEx, O2, Dual CO/H2S, Cl2

This Drager X-am 5000 features a combustible gas Ex sensor, an O2 oxygen sensor, a combined carbon monoxide and hydrogen sulphide sensor and also a chlorine sensor.  It’s ideal for personal monitoring applications and use in explosive atmospheres.

It is a multi-gas detector with the ability to monitor up to five gases.  The tough rubber housing shock-proof sensors make it a robust and reliable instrument.

The Ex sensor in particular has a high level of sensitivity and resistance against sensor poisons such as silicone and hydrogen sulphide.

The X-zone 5000 can be used in alongside the X-am 5000 for area monitoring applications and the Drager Bump Test Station and X-dock testing and calibration station can both be used to carry out functional tests and bump tests.

Key Features

  • Resistant to electromagnetic interference
  • External pump available for pre-entry checks into confined spaces
  • Retro-reflective label makes it easy to see
  • Large display
  • Energy save mode increases operating time to > 40 hours
  • Datalogging capability >1000 hours with 5 gases and a recording interval of 1 value per minute
  •  Results can be read out via Infrared
  • Gas inlets are featured at the top and bottom of the device
Specification
Range (ppm)
  • 0 to 100% LEL
  • 0 to 100 Vol. -% CH4
  • 0 to 25 Vol. -% O2
  • 0-2000 ppm CO
  • 0-200 ppm H2S
  • 0-20 ppm Cl2
Resolution (ppm)
  • 1% LEL
  • 0.1 Vol.-% - O2
  • 1 ppm H2S
  • 2 ppm CO
  • 0.05 ppm Cl2
Response Time (T90)
  • 10 sec (Ex and O2)
  • 20 sec (CO & H2S)
  • 30 sec (Cl2)
Dimensions 48 x 130 x 44 mm; 1.89 x 5.12 x 1.73 in.
Weight 220 - 250 g; 7.8 - 8.8 oz.
Ambient Conditions
  • Temperature: -20 - +50 °C
  • Pressure: 700 - 1300 mbar
  • Humidity: 10 - 95% r.h.
Operating Time >12 hours (without Ex sensor > 700 hours), >40 hours – save energy mode
Charging Time  < 4 hrs
IP Rating IP 67 – water and dust resistant
Alarms
  • Visual (360°)
  • Audible (multi-tone >90 dB at 30cm)
  • Vibrating
Certifications 
  • ATEX – I M1 Ex ia I II 1G Ex ia IIC T3 (Zone 0) I M2 Ex d ia I II 2G Ex d ia IIC T4/T3 Measurement performance certificate according to: EN 50104 (2002)+A1(2004) – O2 EN 60079-29-1:2007 – Methane to Nonane EN 50271:2001 – Software and documentation
  • UL – Class I & II, Div 1 Group A, B, C, D, E, F, G T-Code T4/T3
  • CSA – Class I Div. 1 Group A, B, C, D, T.-Code T4/T3
  • IECEx – Ex ia I Ex ia IIC T3 Ex d ia I Ex d ia IIC T4/T3
  • CE Mark – Electromagnetic Compatibility (Directive 2004/108/EG); EN 50270:2006
  • MED – Marine Equipment Directive (Directive 96/98/EC)
  • MSHA
What's Included
  • Drager X-am 5000 Monitor with specified sensors
  • Crocodile clip
  • NiMH rechargeable power pack and charger
  • Operation and Maintenance manual
Downloads
Reviews
Product Questions