Drager Hydrogen Peroxide Sensor (H2O2) LC

PRODUCT CODE:
6809705

Detect high concentrations of hydrogen peroxide (H202) with this electrochemical sensor. Ideal gas detector sensor for Drager fixed gas systems.

£1,022.82 inc VAT
£852.35
Loading...

Product description

Drager Sensor Hydrogen Peroxide (H2O2) LC

This Drager Sensor detects levels of low concentration Hydrogen Peroxide (H2O2) in the ambient air.  It's an electrochemical sensor, suitable for use in Drager's Fixed Gas Detection Systems, including the PointGard 2100. More information can be found in the technical note in our downloads section.

Specification

Default Settings

  • Measured Gas: Hydrogen Peroxide
  • Display: (H2O2)
  • Chem. Symbol: H2O2
  • CAS number: 7722-84-1

Measuring Range Limit

  • Default: 5ppm
  • Adjustment Range Min/Max: 1/300ppm

Calibration Interval

  • Default: 6 months
  • Adjustment Range Min/Max: 1 day/12 months

Warm-up Time 

  • Ready for operation after max.: 30 mins
  • Ready for calibration after max.:720 mins
  • When using SensorReady: <5 minutes

Detection Limit*:

0.1ppm

Measurement Accuracy* (measurement uncertainly (of meas. value) or minimum (whichever is greater value))

  • ≤ ±5 %
  • ≤ ±0.05ppm

Alarm response time * (on gas exposure with 5x alarm threshold, t0...20 with 1.6x alarm threshold, t0...63

  • ≤5 seconds
  • ≤15 seconds

Loss of sensitivity, per year

≤ –3 %

Expected Service Life (in ambient air)

>36 months

Environmental conditions 

  • Temperature, min./max.: –0/65°C (–32/149°F)
  • Rel. humidity, min./max.: 5/95%
  • Ambient pressure: ±3 %

Stored Conditions

Packed (min./max) 0/40°C (32/104°F)

Cross-Sensitivities

Existing - please refer to manufacturer

Downloads
Reviews