TSI High Flow Flowmeter- 200 L/min (Air / O2 / N2) 1/2"
This high flow flowmeter (model 4043) can be used in a number of applications and industries which require gas flow measurement. This particular model has an inlet diameter of 0.5”, a flow range of 0-200 std l/min and can use air, O2, N2, or an Air/O2 mixture of gas.
TSI’s mass flowmeters provide highly accurate results, to of ±2% of a reading. This is possible due to the platinum film sensor, designed for measuring gas flows. The response of the flowmeter is extremely fast and will respond within 4 milliseconds.
For more information please refer to the datasheet.
- Turn-down ratios greater than 1000:1
- Low pressure drop taken into consideration
- LCD display shows l/min and std l/min
Dimensions |
7.2” x 2.5” x 2.1” (182 x 63 x 53 mm) |
Measurement range |
0-200 std l/m |
Accuracies |
Air & O2 - ±2% of reading or 0.05 Std L/min, whichever is greater N2 - ±3% of reading or 0.1 Std L/min, whichever is greater Air & O2 Mixture - ±3% of reading or 0.1 Std L/min, whichever is greater |
Response |
4 ms to 63% of full scale flow |
Volume* Measurement
|
Range - 0.01 – 99.9 liters Accuracy - ±2% of reading (see Operator’s Manual for additional details) |
Pressure Measurement |
Range - 50-199 kPa absolute Accuracy - ±1 kPa Response - <4 ms to 63% of final value for step charge |
Temperature Measurement |
Range - 0-50°C Accuracy - ±1°C at flow greater than 1 Std L/min Response - <75 ms to 63% of final value for step change |
Ouputs |
Analog - 0-10 VDC flow only, zero and span adjustable via RS232 Digital - RS232 |
DC Power Input |
7.5 VDC ±1.5 V, 300 mA max |
Includes |
Power supply, computer cable (mini-DIN to 9-Pin D-Sub), analog cable (mini-Din to tinned-wire), RS232 serial command set manual, operator’s manual, calibration certificate, inlet filter |