Drager Detection Tubes - Diethyl Ether 100/a (Pack of 10)
These Drager short-term detection tubes are for diethyl ether (100/a). They have a measurement range of 100 to 4000 ppm and are able to detect a range of other substances including Dibutylether, Dichlorodiisopropylether, Dimethyl Ether, Dipropylether, Ether , Ethyl Ether , Isopropyl Ether, Methyl-tert-butylether and Vinyl Trimethoxysilane.
Please note that these tubes have a shelf life of 2 years and come in a pack of 10 tubes.
The Drager Gas Detection Pump (Accuro) is compatible with the short-term gas detection tubes. You can purchase this product by clicking here.
- Hard-to-reach places can be accessed and measured using various compatible accessories
- Excellent for on-the-spot measurements
- 10 seconds to 15 minutes is the time frame for readings and is totally dependent on the tube type
- Suitable applications include pre-entry checks, confined space entry, variations in concentration levels and also contamination of breathing areas are amongst a number of potential of applications
Application Range | |
Standard Measuring Range: | 100 to 4,000 ppm |
Number of Strokes n: | 10 |
Time for Measurement: | approx. 3 min |
Standard Deviation: | ± 15 to 20 % |
Color Change: | orange --> green brown |
Ambient Operating Conditions | |
Temperature: | 15 to 40 °C |
Absolute Humidity: | 3 to 15 mg H2O / L |
Reaction Principle | |
C2H5-O-C2H5 + CrVI --> CrIII + various oxidation products | |
Cross Sensitivity | |
Many petroleum hydrocarbons, alcohols, aromatics and esters are indicated as well, but with different sensitivities. It is not possible to differentiate between them. |