Drager Detection Tubes - Oil Mist 1/a (Pack of 10)
These Drager short-term gas detection tubes are for the measurement of oil mist (1/a). They have a measurement range of 1 to 10 mg/m3 and come in a pack of 10 tubes.
These particular tubes have a shelf life of 2 years.
Buy the Drager Accuro Gas Detection Pump by clicking here. This is required for the use of short-term gas detection tube.
- Obtain readings from hard to reach places with the tubes, pump and accessories
- Carry out measurements on the spot
- 10 seconds to 15 minutes is the potential time frame for readings to be taken and is dependent on the type of tube used
- Pre-entry checks and confined space entry are two applications which can use tubes
- Additional applications may include variations in concentration levels and also contamination of breathing areas
Application Range | |
Standard Measuring Range: | 1 to 10 mg/m3 Discoloration compared to color standard. |
Number of Strokes n: | 100 |
Time for Measurement: | approx. 25 min |
Standard Deviation: | ± 30 % |
Color Change: | white --> brown |
Ambient Operating Conditions | |
Temperature: | 10 to 30 °C |
Absolute Humidity: | < 20 mg H2O / L |
Reaction Principle | |
Oil mist + H2SO4 ä brown reaction product | |
Cross Sensitivity | |
The tube indicates only mineral and synthetic oil aerosols (mist). Oil vapors and vapors of other higher molecular weight organic compounds are not indicated. | |
Additional Information | |
After performing the required 100 pump strokes the reagent ampoule must be broken and the liquid carefully drawn onto the indication layer using the pump. |