Drager Hydrogen High Concentration (0-4 Vol%) Sensor

PRODUCT CODE:
6811365
Drager Hydrogen High Concentration (0-4 Vol%) Sensor
£725.82 inc VAT
£604.85
Loading...

Product description

Drager Hydrogen High Concentration (0-4 Vol%) Sensor

Used for the detection of hydrogen in pariculary high concentrations, this DragerSensor XS EC has a measuring range of 0 to 4 vol.%.

The Drager Safety sensor is used with the Drager X-am 7000 multi gas detector and may be applicable in market segments including ammonia synthesis, fuel refinement (hydrocracking), sulfur elimination, chemical, rocket fuel, leakage inspection, metal processing, industrial gases, fertilizer production, battery chargers and fuel cells.

Key Features

  • This is a plug and play sensor which can be easily replaced
  • Comes with a 1 year guarantee
  • Covers entire range of LELs up to 4 Vol % H2
  • Ideal when using IR technology in X-am 7000
  • Offers high selectivity and linearity
Specification

Detection Limit

0.02 Vol %

Resolution

0.01 Vol %

Response Time

≤ 20 seconds at 20 °C or 68 °F (T50)

Measurement accuracy

Zero Point:  ≤ ± 0.05 Vol. %
Sensitivity:  ≤ ± 2% of measured value

Long-term drift, at 20°C (68°F)

Zero Point:  ≤ ± 0.05 Vol. %/year
Sensitivity:  ≤ ± 3% of measured value/month

Warm-up time:

≤ 1 hour

Ambient conditions

Temperature: (–20 to 50)°C (–4 to 122)°F
Humidity: (10 to 90)% RH
Pressure: (700 to 1,300) hPa

Influence of Temperature

Zero Point:  ≤ ± 0.05 Vol. %
Sensitivity:  ≤ ± 5% of measured value

Influence of Humidity

Zero Point:  No Effect
Sensitivity:  ≤ ± 0.1% of measured value/% RH

Test Gas

approx. 2 Vol. % H2 test gas

Reviews
Product Questions